C3050 - C3051 CH.1: pH, mV, Conductivity + Ref. + °C CH.2: pH, mV, Conductivity + Ref. + °C CH.3: pH, mV + Ref. + °C |
MODE=Selects the settings or escapes from error traps, calibration procedures, etc..by returning to the original mode.
↑↓=Buttons for browsing between the selected modes, for entering a value or for selecting a function.
OK=Proceeds a function.
CAL=Starts a calibration.
STORE=Stores the displayed value or send it to a computer or printer.
HOLD=Holds display when measuring.
The measuring electrodes should be connected to the coaxial connectors CH1 or CH2. Automatic temperature compensation and measurement are possible by connecting a Pt1000 temperature compensator to the banana connectors °C1 or °C2. Without compensator, the manual temperature compensation is automatically switched on. Model C34xx has two extra DIN connectors in parallel with CH.1 and CH2. NEVER use DIN- and BNC- channels together!
A standard USB or Ethernet port is provided for interfacing the instrument with a computer or network.
A standard RS232 port is provided for interfacing the instrument with a printer. Data is sent in the ASCII code at a BAUD rate of 1200…115200 bps (8 bit, no parity, 1 stopbit).
Press MODE to go to the menu. Press OK to enter the selected item. Press MODE to go back or cancel.
BUFFER 1 : 7.00pH BUFFER 2 : 4.00pH BUFFER 3 : –.–pH |